Datasheet MAX30102 (Maxim) - 3

ManufacturerMaxim
DescriptionHigh-Sensitivity Pulse Oximeter and Heart-Rate Sensor for Wearable Health
Pages / Page32 / 3 — Electrical Characteristics (continued). PARAMETER. SYMBOL. CONDITIONS. …
File Format / SizePDF / 917 Kb
Document LanguageEnglish

Electrical Characteristics (continued). PARAMETER. SYMBOL. CONDITIONS. MIN. TYP. MAX. UNITS

Electrical Characteristics (continued) PARAMETER SYMBOL CONDITIONS MIN TYP MAX UNITS

Model Line for this Datasheet

Text Version of Document

MAX30102 High-Sensitivity Pulse Oximeter and Heart-Rate Sensor for Wearable Health
Electrical Characteristics (continued)
(VDD = 1.8V, VLED+ = 5.0V, TA = -40°C to +85°C, unless otherwise noted. Typical values are at TA = +25°C) (Note 1)
PARAMETER SYMBOL CONDITIONS MIN TYP MAX UNITS PULSE OXIMETRY/HEART-RATE SENSOR CHARACTERISTICS
ADC Resolution 18 bits Red ADC Count LED1_PA = 0x0C, LED_PW = 0x01, (Note 2) REDC SPO2_SR = 0x05, 65536 Counts ADC_RGE = 0x00 IR ADC Count LED2_PA = 0x0C, LED_PW = 0x01, (Note 2) IRC SPO2_SR = 0x05 65536 Counts ADC_RGE = 0x00 LED1_PA = LED2_PA = 0x00, 30 128 Counts Dark Current Count LED_DCC LED_PW = 0x03, SPO2_SR = 0x01 ADC_RGE = 0x02 0.01 0.05 % of FS ADC counts with finger on Red sensor under direct sunlight LED 2 Counts DC Ambient Light Rejection ALR (100K lux), ADC_RGE = 0x3, LED_PW = 0x03, IR SPO2_SR = 0x01 LED 2 Counts 1.7V < VDD < 2.0V, LED_PW = 0x01, SPO2_SR = 0x05 0.25 1 % of ADC Count—PSRR (V FS DD) PSRRVDD Frequency = DC to 100kHz, 100mVP-P 10 LSB 3.1V < VLED+, < 5.0V, LED1_PA = ADC Count—PSRR LED2_PA = 0x0C, LED_PW = 0x01, 0.05 1 % of FS (LED Driver Outputs) PSRRLED SPO2_SR = 0x05 Frequency = DC to 100kHz, 100mVP-P 10 LSB ADC Clock Frequency CLK 10.32 10.48 10.64 MHz LED_PW = 0x00 69 LED_PW = 0x01 118 ADC Integration Time INT µs LED_PW = 0x02 215 LED_PW = 0x03 411 Slot Timing (Timing Between LED_PW = 0x00 427.1 Sequential Channel Samples; LED_PW = 0x01 524.7 e.g., Red Pulse Rising Edge To INT µs LED_PW = 0x02 720.0 IR Pulse Rising Edge) LED_PW = 0x03 1106.6
COVER GLASS CHARACTERISTICS (Note 3)
Hydrolytic Resistance Class Per DIN ISO 719 HGB 1 www.maximintegrated.com Maxim Integrated │ 3