5.5 Digit LED Bench Multimeter TTi 1906
Main Features:
Detailed Description
- 5½ digits (±210,000 counts), auto/manual range
- 1µV, 1nA, 1mOhm resolution, 0.012% accuracy
- Wide range of computing functions included
- Automatic data logging functions included
- Integral True RMS AC converter
- Isolated addressable RS232 interface standard
- Model with GPIB (IEEE-488) interface available
Specifications
Accuracy |
|||||||
Accuracies apply for 1 year 18°C to 26°C |
|||||||
DC volts |
Range |
Accuracy |
Input Impedance |
Max Input |
|||
200mV |
±0.017% + 3d |
10M or >1G (Hi Z mode) |
1kV DC or |
||||
2V |
±0.012% + 3d |
||||||
20V |
±0.019% + 3d |
10M |
|||||
200V |
±0.019% + 3d |
||||||
1000V |
±0.019% + 3d |
||||||
Resistance |
Range |
Accuracy |
Max. Measuring Current |
Max. Input |
|||
200 |
±0.025% + 4d |
1.5mA |
300V DC or RMS |
||||
2k |
±0.019% + 3d |
1.5mA |
|||||
20k |
±0.019% + 3d |
0.4mA |
|||||
200k |
±0.019% + 3d |
40mA |
|||||
2M |
±0.022% + 3d |
4mA |
|||||
20M |
±0.07% + 3d |
400nA |
|||||
Max open circuit voltage 3.5V. Diode Test on 2k range. |
|||||||
AC volts |
Range |
Accuracy |
|||||
45Hz - 5kHz |
5kHz - 10kHz |
10kHz-20kHz |
20kHz-50kHz |
||||
200mV |
±0.2% + 150d |
±0.2% + 250d |
±3% + 500d |
— |
|||
2V |
±0.2% + 150d |
±0.2% + 250d |
±0.5% + 300d |
±3% + 1500d |
|||
20V |
±0.2% + 150d |
±0.2% + 250d |
±0.5% + 300d |
±2% + 1000d |
|||
200V |
±0.2% + 150d |
±0.2% + 250d |
±0.5% + 500d |
— |
|||
750V |
±0.2% + 150d |
— |
— |
— |
|||
DC Amps |
Range |
|
Accuracy |
Voltage Burden |
Max. Input |
||
200mA |
±0.08% + 12d |
300 mV max |
1A, 300V |
||||
2mA |
±0.08% + 12d |
||||||
20mA |
±0.08% + 12d |
||||||
200mA |
±0.08% + 12d |
||||||
10A |
<2000mA |
±0.08% + 12d |
650mV max |
10A, 300V |
|||
>2000mA |
±0.12% + 12d |
||||||
Additional error at crest factor=3 typically 0.7%, |
|||||||
AC Amps |
Range |
Accuracy 45 Hz - 1 kHz |
Voltage Burden |
Max Input |
|||
200mA |
±0.37% + 100d |
300mV max |
1A, 300V |
||||
2mA |
±0.37% + 100d |
||||||
20mA |
±0.37% + 100d |
||||||
200mA |
±0.37% + 100d |
||||||
10A |
<2000mA |
±0.37% + 100d |
650mV max |
10A, 300V |
|||
>2000mA |
±0.4% + 100d |
||||||
Additional error at crest factor=3 typically 0.7%. |
|||||||
Further specifications |
|||||||
Input |
Input Current |
< 100 pA |
|||||
DC NMR |
> 60dB at 50/60Hz |
||||||
1k Unbalanced CMR |
> 120dB at DC/50Hz/60Hz, DC ranges; |
||||||
Display |
Display Type |
13mm LED, 8 digit |
|||||
Scale Length |
Selectable 5½ digit or 4½ digit |
||||||
Reading Rate |
3 readings/sec (5½ digit). 10 readings/sec (4½ digit) |
||||||
Overrange |
Displays OL if input too great for range |
||||||
Overflow |
Displays OFLO if calculated result overflows display |
||||||
Annunciators |
LEDs for range, function and program modes |
||||||
Computing |
Null |
Operates over full range; values can be stored for every function |
|||||
Digital Filter |
10 options, including optimised default values for each range and function |
||||||
Touch & Hold |
Reading is held when stable |
||||||
% DEV |
Displays % deviation from entered reference value |
||||||
dB |
Displays measurement in dB relative to 1V, 1mA, 1kW or user entered |
||||||
Ax+B |
Linear scaling of results, with offset. |
||||||
Limits |
Reading displayed with H, L, or P (pass) with respect to user-defined |
||||||
Min/Max |
Minimum and maximum reading stored |
||||||
Data Logger |
Manual or automatic storage of 100 measurements, |
||||||
Interfaces |
RS232 |
Baud rates 300, 1200 or 9600. Complies fully with the ARC (Addressable |
|||||
GPIB (Optional) |
Fully complies with IEEE-488.1 and IEEE-488.2 |
||||||
Power |
AC Input |
110/120 Volts AC nominal 50/60Hz or 220/240 Volts AC nominal 50/60Hz |
|||||
Consumption |
25VA max |
||||||
General |
Operating Range |
+5°C to +40°C, 20% to 80% RH |
|||||
Storage Range |
–40°C to +60°C |
||||||
Size |
260(W) × 88(H) × 235(D)mm, excl. handle and feet |
||||||
Weight |